Skip to main content Skip to search Skip to main navigation

SensoGate retractable fitting sets new standards for safety and efficiency, including now in the upstream area

More accurate measured values, less maintenance effort, and lower risk thanks to automation

With up to 80 potential sources of errors, the manual calibration of pH sensors poses the biggest risk for inaccurate measured values and contamination. In the worst case, a batch loss could occur.

In biopharmaceutical production, the calibration and verification of pH sensors occur before and after every batch and so far has usually been done manually. Due to the required disassembly and installation of the sensors, the manual calibration, and documentation, this proves time-consuming, entails substantial personnel expenses, and correspondingly long system downtimes. Furthermore, there is a significant risk that the glass sensor could be damaged during removal, meaning the sensor can no longer be verified at the end of the batch.

As part of a cCare system, the SensoGate WA131P(H) retractable fitting revolutionizes pH measurement in biopharmaceutical upstream processes through the fully automated cleaning, calibration and sterilization (CIP/SIP) of pH sensors.

The retractable fitting has been tested by an independent laboratory in accordance with ISO 15883-3 with respect to cleanability and in accordance with ISO 17665-1 for sterilizability. The results show that the Hygienic by Design feature of the SensoGate retractable fitting makes it the best retractable fitting for the most stringent hygienic requirements.

Information regarding GMP-compliant validation can be derived from the test setup and test results.

SensoGate WA131P(H): the solution for automated pH measuring points

The video shows a short excerpt from the detailed process of cleaning, calibration and sterilization of the sensor in the calibration chamber of the SensoGate WA131P(H) retractable fitting between two batches, as well as a sensor replacement during a running batch.

The SensoGate WA131P(H) supports the following:

  • Fully automated cleaning (CIP), calibration and sterilization (SIP): automated processes reduce the manual effort and eliminate sources of errors.
  • Sensor replacement during running batches: the patented lock-gate system reliably separates the process, prevents media carryover, and makes it possible to replace the sensor without interrupting the process.
  • Meeting the highest hygienic requirements thanks to Hygienic by Design: validated in accordance with ISO 15883-3 (cleanability) and ISO 17665-1 (sterilizability), the fitting meets the highest hygienic requirements.

Use the form to access the full video.

Schedule an appointment for a personal consultation:

Call us: +49 30 264789640
Email us: Contact
Available from: Monday - Friday 8:00 am to 4:00 pm Germany CET/CEST (UTC+1/UTC+2)

Or find your local partner.

SensoGate WA131P White Paper

Request white paper: How SensoGate meets hygienic requirements in upstream processes

Sensor replacement during an ongoing process and automated calibration offer significant potential for greater efficiency in pharmaceutical and biotech facilities. This requires a retractable fitting whose cleanability meets the high hygienic requirements.

This white paper shows how SensoGate meets these requirements:

  • with insights into the certification test,
  • explanations regarding relevant regulations,
  • and specific information regarding the evaluation of hygienic suitability.

Facility planners and QA managers will learn why SensoGate can be used in upstream processes and how its proven hygiene standard can reduce the time and effort involved in validation.

Fill out the form to receive the white paper directly via email.

What challenges do you face during your validation processes?

Fill out the form to access the complete video and white paper, which presents the cleaning, calibration and sterilization process for a sensor, as well as additional detailed technical information and validation documents.

Technical depth: gain insights into the specific design features and performance data of the SensoGate WA131P(H), which go beyond publicly available information.

Validation support: use the detailed test results and the resulting recommendations as the basis for your own validation. This information offers significant added value for your quality control and quality assurance by simplifying and accelerating the validation process.

Secure your own exclusive access:

Fields marked with asterisks (*) are required.

Consent Declaration: I have taken note of the privacy policyand agree to it. I consent to the data I have provided being used by Knick GmbH for processing my inquiry and for contacting me. The whitepaper will only be provided following review and approval by Knick. There is no entitlement to receive the whitepaper or the video. This consent may be withdrawn at any time with future effect by sending an email to [email protected].*


Frequently asked questions (FAQ)

How does a retractable fitting work?

The SensoGate retractable fitting is a high-end system that retracts the pH sensor from the process medium into an integrated rinsing and calibration chamber. In this service position, the sensor is fully automatically cleaned, calibrated, and replaced if necessary, without interrupting the ongoing process. Once these maintenance processes are complete, the sensor is sterilized and returned back into the process position.


What is the advantage of a retractable fitting compared to a static inline fitting?

Compared to static inline fittings, the SensoGate retractable fitting makes it possible to perform maintenance work, such as cleaning, calibration and sensor replacement, without interrupting the process. This reduces downtime and minimizes the risk of contamination, since the sensor is retracted into a separate chamber where the maintenance is performed.


What does Hygienic by Design mean?

Hygienic by Design means that the SensoGate retractable fitting was specifically designed for hygienic applications. Its smooth design that is resilient to dirt, with no dead spaces, allows for complete cleaning and sterilization. All parts that come into contact with the process are made from FDA-compliant materials and are suitable for CIP (Cleaning in Place) and SIP (Sterilization in Place). The rinsing system and cavities are optimized to create a sufficient flow that effectively washes away dirt particles.


How does the cyclone rinsing work?

The cyclone rinsing of the SensoGate retractable fitting works as follows:

  1. The sensor is pneumatically retracted into the service position.
  2. Cleaning medium flows over tangential inlets into the calibration chamber.
  3. A rotating vortex is created that thoroughly cleans the sensor cap, the sensor membrane, the sealing areas, and all of the walls of the chamber.
  4. At the same time, the secondary flow cleans the seal of the immersion tube in order to remove deposits.

This process ensures effective cleaning without leaving inaccessible areas, niches or residue behind.


How does an automatic cleaning process work?

The automatic cleaning process is performed through the combination of the SensoGate retractable fitting with the cCare system. During the process, the sensor is retracted into the service position where it is thoroughly cleaned by the cyclone rinsing. An automatic calibration can then be performed before the sensor is introduced back into the process.


How does an automatic calibration work?

The automatic calibration is performed by the cCare system, which calibrates the sensor with buffer solutions after cleaning in the calibration chamber. This process is fully automated and standardized, increasing the accuracy and reproducibility of the measurements.


How does a sensor replacement work while the process is ongoing?

A sensor replacement while the process is ongoing is possible thanks to the patented lock-gate system of the SensoGate retractable fitting. The sensor is retracted into the service position where it can be replaced without interrupting the process. The process is securely separated. Before the sensor is moved back into the process, it and all surfaces that come into contact with the process are sterilized to eliminate the risk of contamination.


Why is automatic calibration better than manual calibration?

Automatic calibration offers several advantages compared to manual calibration:

  • Reduction of sources of error: automated processes minimize human errors during handling and documentation.
  • Time and cost savings: automatic calibrations reduce personnel costs and system downtime.
  • Increased measurement accuracy: standardized calibration processes ensure consistent and precise measurement results.


What risks and possible errors are minimized by the automatic calibration?

The following risks and possible errors are minimized by the automatic calibration:

  • Handling errors: incorrect operations during removal, cleaning or sensor installation.
  • Documentation gaps: incomplete or erroneous calibration records.
  • Contamination risks: ingress of contaminants during manual handling.
  • Measurement errors: inaccurate calibrations lead to erroneous measured values.


How quickly does a cCare system pay for itself?

The payback period for a cCare system depends on different factors, such as the frequency of calibrations, personnel costs, and the cost of production downtime. However, the system usually pays for itself within a few years due to the maintenance savings and the increased production times.


How much time between two batches does an automatic maintenance system save?

An automatic maintenance system like cCare can significantly shorten the time between two batches, since there is no manual calibration or cleaning. The exact time saved depends on the specific process conditions, but can be significant.

The calibration is very time-consuming and requires an employee to “enter” the cleanroom, remove the sensor, clean it, swirl it in various buffer solutions one after the other, and wait until the pH value is stable.

An automatic maintenance system like cCare can significantly shorten the time between two batches, since the time is eliminated for the employee to enter the cleanroom and perform the maintenance, which involves removing the sensor, cleaning it, swirling it in various buffer solutions one after the other, and waiting until the pH value is stable.

Furthermore, the sensor maintenance can then be performed once the batch is complete, rather than having to wait for the maintenance staff.

The time savings depend on the specific process conditions, but can be significant and increase production efficiency.